ZEISS IOLMaster 700
Roll over image to zoom in
€ 43.000,00
The IOLMaster 700 device is a combined biometry instrument for the visualization of eye structures and acquisition of data of the human eye required for the calculation of the intraocular lens to be implanted.
Specifications
Measurement range
|
Axial length
|
14 – 38 mm
|
---|---|---|
Corneal radii
|
5 – 11 mm
|
|
Anterior chamber depth
|
0.7 – 8 mm
|
|
Lens thickness
|
1 – 10 mm (phakic eye)
|
|
0.13 – 2.5 mm (pseudophakic eye)
|
||
Central corneal thickness
|
0.2 – 1.2 mm
|
|
White-to-white
|
8 – 16 mm
|
|
SD of repeatability
|
Axial length
|
5 μm
|
Corneal radii
|
0.09 D
|
|
Cylinder
|
> 0.75 D, axis 3.8°
|
|
Anterior chamber depth
|
7 μm
|
|
Lens thickness
|
6 µm
|
|
Central corneal thickness
|
2.5 µm
|
|
White-to-white
|
111 µm
|
|
IOL calculation formulas
|
Barrett Suite: Barrett Universal II & Barrett TK Universal II*, Barrett toric & Barrett TK toric*, Barrett True K & Barrett True K with TK
|
|
Haigis Suite: Haigis, Haigis-L, Haigis-T
|
||
Hoffer® Q |
||
Holladay 1 and 2
|
||
SRK®/T |
||
Interfaces
|
ZEISS EQ Workplace
|
|
ZEISS VERACITY Surgery Planner |
||
ZEISS EQ Mobile
|
||
ZEISS FORUM eye care data management system
|
||
ZEISS computer-assisted cataract surgery system CALLISTO eye (via USB & FORUM)
|
||
Data interface for electronic medical record (EMR) / patient management systems (PMS), Holladay IOL
|
||
Consultant software and PhacoOptics® |
||
Data export to USB storage media
|
||
Ethernet port for network connection and network printer
|